Services Available | |
---|---|
Repair | Yes |
Calibration | Yes |
Free Support | Yes |
El LP02, fabricado por Hukseflux, es un piranómetro "second class" ISO 9060:2018 Class C, que mide la radiación solar global. La salida es en mV, de forma que directamente la puede leer un datalogger Campbell Scientific. Este piranómetro se puede utilizar en variedad de apliaciones meteorológicas.
Leer másThe LP02 measures solar radiation with a high-quality blackened thermopile protected by a dome. The blackened thermopile provides a flat spectral response for the full solar spectrum range, which enables the LP02 to be used under plant canopies or lamps, when the sky is cloudy, and for reflected radiation measurements.
The LP02 includes a bubble level, three adjusting screws, and a cable gland. The bubble level and adjusting screws allow the sensor to be leveled without using a leveling base. The gland facilitates cable replacement.
The LP02 produces a millivolt signal that is measured directly by a Campbell Scientific datalogger.
Two LP02 pyranometers can be mounted back-to-back to make a low-cost albedometer; contact Campbell Scientific for more information.
Número de FAQs relacionadas con LP02: 3
Expandir todoDesplegar todo
Most Campbell Scientific sensors are available as an –L, which indicates a user-specified cable length. If a sensor is listed as an –LX model (where “X” is some other character), that sensor’s cable has a user-specified length, but it terminates with a specific connector for a unique system:
If a sensor does not have an –L or other –LX designation after the main model number, the sensor has a set cable length. The cable length is listed at the end of the Description field in the product’s Ordering information. For example, the 034B-ET model has a description of “Met One Wind Set for ET Station, 67 inch Cable.” Products with a set cable length terminate, as a default, with pigtails.
If a cable terminates with a special connector for a unique system, the end of the model number designates which system. For example, the 034B-ET model designates the sensor as a 034B for an ET107 system.
Not every sensor has different cable termination options. The options available for a particular sensor can be checked by looking in two places in the Ordering information area of the sensor product page:
If a sensor is offered in an –ET, –ETM, –LC, –LQ, or –QD version, that option’s availability is reflected in the sensor model number. For example, the 034B is offered as the 034B-ET, 034B-ETM, 034B-LC, 034B-LQ, and 034B-QD.
All of the other cable termination options, if available, are listed on the Ordering information area of the sensor product page under “Cable Termination Options.” For example, the 034B-L Wind Set is offered with the –CWS, –PT, and –PW options, as shown in the Ordering information area of the 034B-L product page.
Note: As newer products are added to our inventory, typically, we will list multiple cable termination options under a single sensor model rather than creating multiple model numbers. For example, the HC2S3-L has a –C cable termination option for connecting it to a CS110 instead of offering an HC2S3-LC model.
Many Campbell Scientific sensors are available with different cable termination options. These options include the following:
Note: The availability of cable termination options varies by sensor. For example, sensors may have none, two, or several options to choose from. If a desired option is not listed for a specific sensor, contact Campbell Scientific for assistance.
Nota: lo siguiente muestra información de compatibilidad notable. No es una lista de todos los productos compatibles.
Producto | Compatible | Nota |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR300 (retired) | ||
CR3000 (retired) | ||
CR310 | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) | ||
CR9000X (retired) |
The LP02 includes a bubble level and three adjusting leveling screws, which allow the sensor to be leveled without using a leveling base. The CM225 Solar Sensor Mounting Stand is used to attach the sensor to a crossarm. The CM225 consists of a rectangular plate, mounting bracket, U-bolts, washers, lock washers, and nuts. The LP02 should be mounted away from all obstructions and reflective surfaces that might adversely affect the measurement.
If the sensor needs to be mounted at an angle, a CM255 or CM255LS adjustable-angle mount can be used instead.
Sensor | Blackened thermopile protected by a dome |
Measurement Description | Monitors solar radiation for the full solar spectrum range |
Spectral Range | 285 to 3000 nm |
Maximum Irradiance | 2000 W/m2 |
Sensitivity | 15 µV/W/m2 (nominal) |
Operating Temperature Range | -40° to +80°C |
Temperature Dependence | < 0.15% per °C |
ISO Classification | ISO 9060:2018 spectrally flat Class C (second class) |
Body Diameter | 7.8 cm (3.1 in.) |
Dome Diameter | 3.0 cm (1.2 in.) |
Height | 5.9 cm (2.3 in.) |