What might cause a CR23X to lock up and display an "E08 Watchdog Reset" error?

E08 errors are occasionally created by voltage surges or transient voltages. Section 3.9 of the CR23X manual explains in greater detail what a watchdog is and possible causes.  If these are happening frequently, it is likely an issue with the system  that needs to be resolved. The first step in this resolution process would be to check the system power (inputs and outputs) and verify it is within the expected tolerances as outlined in the Specifications page (OV-23) of the CR23X Micrologger Operator's Manual