HMP155A-L Sensor temperatura y humedad relativa del aire
Accurate, Wide Temp Range
Higher-end sensor where higher accuracy is required
meteorología aplicaciones agua aplicaciones energía aplicaciones flujo gas y turbulencia aplicaciones infraestructuras aplicaciones suelo aplicaciones

Resumen

The HMP155A, manufactured by Vaisala, monitors relative humidity (RH) for the range of 0 to 100% RH and temperature for the range of -80° to +60°C. It can provide reliable measurements for a wide range of applications, as part of a weather station system or as a single instrument. All Campbell Scientific dataloggers are compatible.

Leer más

Ventajas y características

  • Adecuado para aplicaciones desatendidas y a largo plazo
  • Precisa y robusta
  • Se monta en un mástil de trípode o torreta, brazo, o poste suministrado por el usuario
  • Compatible con la mayoría de dataloggers Campbell Scientific

Imágenes

Descripción detallada

The HMP155A uses a HUMICAP180R capacitive thin film polymer sensor to measure RH, and a PRT to measure temperature. 

A warmed probe option ensures accurate measurements and better long-term stability in high-condensing environments, where required.

To reduce the current drain, power can be supplied to the HMP155A only during measurement when the sensor is connected to the datalogger’s switched 12 V terminal. Dataloggers that do not have a switched 12 V terminal, such as the CR510 or CR7, can use the SW12V switched 12 V device to switch power to the sensor only during measurement.


Preguntas frecuentes

Número de FAQs relacionadas con HMP155A-L: 3

Expandir todoDesplegar todo

  1. Note the difference between calibration and a field check. Calibration cannot be done in the field, as it requires an experienced technician and specialized equipment.

    Field checks of measurements can be done to determine if the data make sense with the real-world conditions. Follow these steps to field check a sensor:

    1. Find a second sensor of the same type as the installed sensor whose data is in question. The second sensor will be used as a benchmark sensor and should be known to be accurate or recently calibrated.
    2. At the site, take readings using both sensors under the same conditions. The best practice is to measure both sensors side-by-side at the same time. Note that the sensors will never have the exact same measurement.
    3. Depending on the sensor model, if the difference in the readings of the installed and benchmark sensors is greater than the sum of the accuracies for both sensors, either return the installed sensor to Campbell Scientific for calibration or replace the appropriate chip.
      • The 107, 108, 109, 110PV-L, and BlackGlobe-L temperature sensors can be calibrated.
      • The HC2S3-L and HMP155A-L temperature and relative humidity sensors can be calibrated.
      • The CS215-L has a replaceable chip for temperature and relative humidity. For more information, refer to the “Maintenance and Calibration” section of the CS215 instruction manual.
      • The HMP60-L has a replaceable chip for relative humidity only. For more information, refer to the “Maintenance” section of the HMP60 instruction manual.
  2. Single-ended channels are typically used when it is desirable to use fewer channels and the cable length is shorter—less than 6.1 m (20 ft). Differential channels are preferable when the cable needs to be longer than 6.1 m (20 ft).

Compatibilidad

Nota: lo siguiente muestra información de compatibilidad notable. No es una lista de todos los productos compatibles.

Dataloggers

Producto Compatible Nota
CR1000 (retired)
CR1000X (retired)
CR300 (retired)
CR3000 (retired)
CR310
CR350
CR6
CR800 (retired)
CR850 (retired)

Información de compatibilidad adicional

Sensor Mounts

When exposed to sunlight, the HMP155A must be housed in a 41005-5 or RAD14 14-plate naturally aspirated radiation shield. The 41005-5 and RAD14 attach to a crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 to 2.1 in.) outer diameter. The RAD14 uses a double-louvered design that offers improved sensor protection from driving rain, snow, and insect intrusion, It also has lower self-heating in bright sunlight combined with higher temperatures (> 24°C [~75°F])
and low wind speeds (< 2 m s-1 [~4.5 mph]), giving a better measurement.

Especificaciones

Electromagnetic Compatibility Complies with EMC standard EN61326-1 Electromagnetic
Filter Description Sintered PTFE
Housing Body Material PC
Housing Classification IP66
Voltage Output 0 to 1 Vdc
Average Current Consumption ≤ 3 mA (analog output mode)
Operating Voltage 7 to 28 Vdc
Settling Time 2 s (at power up)
Field-Replaceable Chip or Recalibrate Recalibrate
Tip Diameter 1.2 cm (0.5 in.)
Length 27.9 cm (11 in.)
Head Height 4 cm (1.6 in.)
Body Height 2.4 cm (0.9 in.)
Body Width 2.0 cm (0.8 in.)

Relative Humidity

Sensing Element HUMICAP 180R
Measurement Range 0 to 100% RH (non-condensing)
Response Time
  • The response time for the RH specification is for the HUMICAP 180R at 20°C in still air with sintered PTFE filter and a 0 to 75% RH step change.
  • 20 s (63% step change)
  • 60 s (90% step change)
Factory Calibration Uncertainty
  • The factory calibration uncertainty is defined as ±2 standard deviation limits. Uncertainty is at +20°C. Small variations are possible; see also the calibration certificate.
  • ±0.6% RH 0 to 40% RH
  • ±1.0% RH 40 to 97% RH
Accuracy
  • -NOTE- Accuracy specifications include non-linearity, hysteresis, and repeatability.
  • ±1% RH (at 15° to 25°C, 0 to 90% RH)
  • ±1.7% RH (at 15° to 25°C, 90 to 100% RH)
  • ±(1.4 + 0.032 × reading) % RH (at -60° to -40°C)
  • ±(1.2 + 0.012 × reading) % RH (at -40° to -20°C)
  • ±(1.0 + 0.008 × reading) % RH (at -20° to +40°C)
  • ± (1.2 + 0.012 × reading) % RH (at 40° to 60°C)

Air Temperature

Sensing Element PT 100 RTD 1/3 class B IEC 751
Measurement Range -80° to +60°C
Accuracy
  • ±(0.226 - 0.0028 x temperature)°C (-80° to +20°C)
  • ±(0.055 + 0.0057 x temperature)°C (+20° to +60°C)
Accuracy (RS-485) ±(0.176 - 0.0028 x temperature) °C (-80° to +20°C)
±(0.07 + 0.0025 x temperature) °C (+20° to +60°C)
Entire Temperature Range Refer to graph in probe manual.

Descargas

HMP155A Example Programs (4 KB) 15-10-2021

Description: Example CR1000X programs that measure either the analog voltage or RS-485 output of the HMP155A. The analog voltage programs use either the VoltSE or VoltDiff instruction and switch power to the sensor only during measurements. One program also calculates the mean vapor pressure. The RS-485 program measures the RS-485 output with an SDM-SIO1A/SIO4A.  Although these programs are written for the CR1000X, programs for other dataloggers will be similar.


Casos de aplicación

United Kingdom: Pioneering Research in Carbon Sequestration
Overview In the fight against climate change, innovative solutions are emerging to address the global challenge......leer más
Zambia: Strengthening Climate Resilience
Overview In May 2019, the Government of Zambia embarked on the Climate Adaptation Water and Energy......leer más
France: Restoring a Royal Chapel
Overview The Royal Chapel of Chaalis Abbey—often referred to as the “French Sistine Chapel”—is a treasure......leer más
Zimbabwe: Enhancing Climate Resilience
Overview As part of the United Nations Development Programme's (UNDP) Climate Adaptation Water and Energy Programme......leer más
Eswatini: Strengthening Early Warning Systems and Climate Services
Overview The Kingdom of Eswatini in South Africa—with its agricultural-based economy and a population of less......leer más
Argentina: Measuring Mountain Weather Conditions
Introduction The Andes Mountains of Argentina are home to a rich and diverse group of organic......leer más