SW12V 12 V Power Switch Assembly

Resumen

The SW12V is a power switch that enables your data logger to determine whether 12 Vdc power is passed onto an external device.

Leer más

Ventajas y características

  • Permite a un datalogger antiguo alimentar un sensor HMP45C o HMP155A sólo durante la medida, y así reducir el consumo de energía

Imágenes

Productos similares

Descripción detallada

The SW12 provides similar functionality as a data logger's switched 12 V terminal. Therefore, this power switch is typically used with older data loggers (e.g., CR510, 21X, CR7) that do not have a switched 12 V terminal.

The SW12V consists of a cable that connects to the data logger, and a terminal block for connecting the external device.

The SW12V has a shrink wrap that lists the function of each wire by color. Wiring to the power source should be red to positive (12 V), black to ground, and green to control (on a Campbell Scientific data logger this is a control port labeled with a “C” or a universal port labeled with a "U."). Power out (to the device being powered) requires a 12 V connection and a ground connection.

Power is distributed to the remote device when the control wire is activated. Wire the power (+ lead) from the device being powered to a connection labeled “SW12V” on the SW12V green connector. Wire the ground (─  lead) to the G terminal on the SW12V green connector.

Especificaciones

Function Enables a data logger to determine whether 12 Vdc power is passed onto an external device.
Current Ratings
  • 0.9 A (@ 20°C)
  • 0.65 A (@ 50°C)
  • 0.36 A (@ 85°C)
Control Line Voltage Range 4 to 16 Vdc
Cable Length 17.7 cm (18 in.)
Weight 45.4 g (0.1 lb)

Compatibilidad

Nota: lo siguiente muestra información de compatibilidad notable. No es una lista de todos los productos compatibles.

Dataloggers

Producto Compatible Nota
CR1000 (retired)
CR200X (retired)
CR216X (retired)
CR3000 (retired)
CR5000 (retired)
CR800 (retired)
CR850 (retired)
CR9000X (retired)

Miscelaneo

Producto Compatible Nota
21X (retired)
CR10 (retired)
CR10X (retired)
CR23X (retired)
CR500 (retired)
CR510 (retired)
CR9000 (retired)

Información de compatibilidad adicional

Sensors

The SW12V is primarily used with the redesigned HMP45C relative humidity probe (effective May, 2004) and the HMP155A relative humidity probe.

Preguntas frecuentes

Número de FAQs relacionadas con SW12V: 4

Expandir todoDesplegar todo

  1. Yes, this is possible. Keep in mind, however, that a common ground between related devices is essential for obtaining a good measurement.

  2. The maximum voltage sourced through the SW12V is 16 Vdc.

  3. The maximum current is dependent on temperature:

    • 0.9 A @ 20° C
    • 0.65 A @ 50° C
    • 0.36 A @ 85° C

    These limits are set by the polyfuse incorporated in the device.

  4. The control line voltage range is 4 to 16 Vdc. A drive current of < 0.5 mA is required.